Features |
|
‧1/16,000 display resolution (Max. 1/60,000 depending on load cell |
|
quality & performance). |
|
‧Internal Resolution 1,000,000, A/D Conversion rate 120 times/ Sec.
‧Watchdog virtually eliminates malfunctions that associated with |
|
computerized equipment or software failure. |
|
‧Full Digital Calibration makes setting ZERO and SPAN Calibration an |
|
easy task. |
|
‧Drives up to 8 parallel connecting load cells. (350 Ω)
‧8k bytes SRAM with Li-battery backup. Information will not appear |
|
even power failure. |
|
 |
| |
Analog input and A/D conversion |
| |
|
Type |
JIF-2002 |
|
Input Sensitivity |
≧0.3μV/D |
|
ZERO Adjustment |
0 ~ 20 mV |
|
Max. Resolution |
|
1/16,000 |
|
(Max. 1/60,000 depending on load cell
quality & performance) |
|
|
Main Display (Green Tube) |
|
7-segment, 7-digit display, |
|
VFD screen with a 13mm character size,
display the weight. |
|
|
Minimum Division |
×1 , ×2 , ×5 , ×10 , ×20 , ×50 |
|
Maximum Display |
+800450 |
|
Load Cell Excitation |
DC10V±1% , 230mA, Remote Sensing. Can be
connected up to 8-(350Ω Load Cells) |
|
Non-Linearity |
0.01% F.S. |
|
A/D Conversion Method |
△Σ |
|
A/D Resolution |
≒1/1,000,000 |
|
A/D Conversion Rate |
Approx. 120 Times / Sec. |
|
Max. Load Cell Input Voltage |
32mV |
|
ZERO Temperature Comp. |
±(10.2μV + 0.001% of Dead Load) /
℃
TYP |
|
SPAN Temperature Comp. |
±0.001%
℃